MANUAL DE INSTRUCCIONES
J
.P. SELECTA s.a.u.
CODIGO 80350 REV B
6 Equipment description
Presoclave III autoclave belongs to a series of saturated vapour autoclaves that
are distinguished for the following features:
1. Arm / joist locking system.
2. Temperature and time electronic control circuit.
3. Safety locking system that only allows lid opening with pressure 0 and temperature
under 95ºC.
4. Maximum pressure limiter that disconnects the heating element in case of over-
pressure.
5. Limiter thermostat to protect the heating element.
6.1 Equipment
-
Safety system:
Safety over pressure valve.
Safety pressure switch.
Safety thermostat.
Lid closed detector.
Protective thermal lid cover.
-
Chamber and lid made in stainless steel.
-
External case made in stainless steel.
-
Manual emptying valve.
-
Waste water tank connection.
6.2 Features
-
Sterilization temperature from 115°C to 134°C.
-
Temperature resolution 1ºC.
-
Sterilization time from 3 minutes to 99 minutes.
-
Time resolution 1 minute.
-
Suitable to work with solids and liquids atmospheric purge from 115°C to 134°C.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
FEBRERO 2020
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 20