Operation; Opening The Lid - Selecta PRESOCLAVE III 50L Manual De Instrucciones

Autoclaves para esterilización a vapor
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL
15
J
.P. SELECTA s.a.u.
CODE 80350 REV B
¡WARNING! DON'T TRY TO OPEN THE LID

8 Operation

8.1 Opening the lid

-
The autoclave must be connected to mains supply and the main
switch turned on.
-
Pressure gauge: 0 bar.
-
Temperature in the internal chamber must be under 95ºC.
-
Press the stop button (12) to unlock the locking system of the lid
opening lever.
ONLY WITH THIS CONDITIONS, LID IS READY TO BE OPEN
8.2 Start
-
Start on using the main switch (17)
-
Open the lid as we can see on 8.1 section, using the handle (16)
-
Place the shelf at the bottom of chamber.
16
-
Be sure the valve (15) is closed.
-
Fill of decalcified water until shelf level.
-
Put the material to sterilize and close the lid.
-
Start the cicle as shown on section 8.3
-
When closing the lid, make sure that the open door indicator (2)
is off, otherwise the device will not start.
BEFORE STARTING THE CICLE, ALWAYS
CHECK IF THERE IS ENOUGH WATER IN THE
CHAMBER. IF NOT, THE HEATING ELEMENT
COULD BE DAMAGED.
Autovía A2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: www.jpselecta.es
FEBRUARY 2020
UNLESS READING
«8 OPERATION» SECTION
Page: 23
Tel 34 937 700 877 Fax 34 937 702 362

Hide quick links:

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Presoclave iii 80l40017584001759

Tabla de contenido