7. What do I do if...
... no weight is displayed with a load on the scale?
– Is the scale switched on?
– Check the mains unit connection.
0. 0 00
...
does not appear before weighing?
– Press the Start key again - there must not be any load on the scale.
... one segment is illuminated permanently or not at all?
– That point has a fault. Inform your service agent.
... the display shows
– Battery voltage is dropping. Charge the rechargeable batteries in the next few days.
batt
...
appears in the display?
– The rechargeable battery is so severely discharged that weighing is no longer possible.
Recharge the battery.
St0p
...
appears in the display?
– Maximum load has been exceeded.
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
Er:011:
...
appears in the display?
– The scale is overloaded or the load is unevenly distributed. Take the load off the scale
or distribute weight more evenly and restart scale.
Er:012:
...
appears in the display?
– The scale was switched on with too great a load on it. Remove load and restart.
8. Maintenance/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 18).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
Model 336
?
appears?
19
GB