ELECTROMAGNETIC COMPATIBILITY
Guidance and manufacture's declaration - electromagnetic immunity
This monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of this monitor should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic discharge
± 6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
Voltage dips, short
<5 % UT
interruptions and
(>95 % dip in UT)
voltage variations
for 0.5 cycle
on power supply
40 % UT
input lines
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
36 • ENGLISH
For all ME EQUIPMENT and ME SYSTEMS
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
Table 2
Compliance Level
Electromagnetic Environment Guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
this monitor requires continued operation during
power mains interruptions, it is recommended that
this monitor be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.