Troubleshooting (cont.)
If the door is fully raised for more than 20 minutes when there is no water in the bath,
the seals will automatically deflate.
It is normal to hear the solenoid valve, located inside the door, open and expel air when
the water level falls below the sensors. The valve is deflating the door seal.
Water coming from the bottom of the door and entering the receptor tray is normal when
it is caused by the bather moving about in the bath when it is completely full. Excessive
splashing will cause water to run over the top of the inflated seals into the receptor tray.
Water dripping into the receptor tray as the seals deflate is normal. Water collects on the
seal during use and falls into the receptor tray when the seals deflate.
Troubleshooting the Bath System
Symptoms
1.
Door is difficult to raise or
lower.
2.
Door shimmies or shakes.
3.
Water is leaking/running
from accumulator.
Kohler Co.
Probable Causes
A. One or both counterweights are
not attached.
B. One or both of the
counterweight cables are not fed
through the pulleys.
C. The bath was not leveled
properly, causing
counterweights to rub or
generate a noise.
D. Mounting bracket is not
adjusted properly.
E. Door is rubbing against the bath
or access panel.
F.
A roller is damaged or has
fallen off.
G. There is excessive wear on the
latch pin and/or strike plate.
A. Mounting bracket needs
adjustment.
A. Normal operation. The
accumulator is draining into the
receptor.
B. Water supply connection is
loose or damaged.
C. Accumulator fitting(s) are loose
or disconnected.
39
Recommended Action
A. Reattach the counterweight(s).
B. Use the provided tool to run the
counterweight cable onto the
pulleys.
C1. Level the bath.
C2. Confirm the counterweights do not
contact the receptor when the door
is up. If there is contact the
receptor drain is not low enough in
the subfloor. Enlarge the drain
cutout and lower the receptor or
raise the bath using the adjustable
feet.
D. Reduce the pressure on the door by
adjusting the mounting brackets
outward using the slots.
E1. Adjust the door outward using the
slots in the pulley assembly.
E2. Adjust the lower access panel
bracket rearward using the slots.
E3. Adjust the brush seal under the
door forward using the slots.
F.
Replace the roller.
G1. Apply a drop or two of oil to the
face of the strike plate.
G2. Adjust the rubber bumper to center
the latch pin in the strike plate
opening when the door is raised.
A. Adjust one or both of the mounting
brackets inward using the slots,
until it is snug against the rollers.
A. See the "Normal Operation
Includes" portion of the
"Troubleshooting."
B. Reseal the water inlet connections
and/or reconnect them.
C. Replace the accumulator connection
fitting(s).
1128532-2-H