Troubleshooting

5
Troubleshooting
Fault
The words "Remote System"
are not displayed on the
analyzer.
Measured value too low
The display goes out during
the flushing phase.
Recurring pump errors
® DrägerService is a registered trademark of Dräger.
6

Construction and function

The pump of the Remote System continuously draws in the sample gas through the sampling
hose and gas inlet. After a substance-specific flushing time, the analyzer draws in the gas
sample quantity required for measurement.
Sample gas
26
Cause
The pump is not fully screwed
in place.
Flushing time too short
Batteries discharged
Gas inlet on the pump
clogged.
Analyzer blocked
Pump
Outlet
Remedy
Check the connection and the
electrical plug-in
connections.
Call DrägerService
Wait for the complete flushing
time.
Insert new batteries.
Call DrägerService
to chip
Analyzer
Dräger CMS Remote System
®
.
®
.