6. Error messages

LCD Display
Area
Measured
temperature
too high
Measured
temperature
too low
Ambient
temperature
too high
Ambient
temperature
too low
Error display
(0 & 6)
Error function
display 2
Low battery
indicator
Description
Displays «H» when
measured temperature is
higher than 109.4 °F / 43
°C in body mode or 211.8
°F / 99.9 °C in object
mode.
Displays «L» when
measured temperature is
lower than 93.2 °F (34.0 °C)
in body mode or 32.2 °F
(0.1 °C ) in object mode.
Displays «AH» when
ambient temperature is
higher than 104.0 °F
(40.0 °C).
Display «AL» when
ambient temperature is
lower than 59.0 °F (15.0
°C) in body mode or
lower than 41.0 °F (5.0
°C) in object mode.
System malfunction.
The sensor is too close or
touching object. Move
the sensor 2" away from
the object (5 cm). Do not
touch sensor.
If the steady battery icon
is the only symbol shown
on the display, the batter-
ies should be replaced
immediately.