Descargar Imprimir esta página

Lovibond Tintometer MD 100 Manual De Instrucciones página 11

Ocultar thumbs Ver también para Tintometer MD 100:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 25
EN
Add DPD No. 1 tablet .
Fill up vial with sample to
the 10 ml mark.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/l Ozone appears on the display.
12
Ozone T / M300
Add DPD No. 3 tablet .
Close vial(s).
Test
Press the TEST (XD:
START)button.
24 mm
Crush tablet(s) by rotating
slightly.
Dissolve tablet(s) by
inverting.
12
11
10
9
8
7
6
Wait for 2 minute(s)
reaction time.
EN   Method Reference Book   1.0
2 min
1
2
3
4
5

Publicidad

loading

Este manual también es adecuado para:

Tintometer md 110Tintometer md 200