Phcbi MCO-230AIC Instrucciones De Operación página 65

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Alarm or safety
function
(MCO-230AICUV, or when optional
UV lamp failure
MCO-170UVS is installed)
The UV lamp burns out.
(MCO-230AICUV, or when optional
New UV lamp
MCO-170UVS is installed)
replacement
The accumulated ON time reaches
approx. 5,000 hours.
Communication between LCD touch
Communication
panel and control substrate is
error
terminated or unstable.
After power switch is turned ON, under
Warming-up of
warming-up before temperature is
gas control
stable and gas control is enabled.
Tables 5–7 show the behaviour of the alarm (buzzer) and ring back function when pressing the Buzzer key.
Table 5. In cases other than those covered in Tables 6 and 7
Remote Alarm setting
ON: Remote alarm setting not connected with
Buzzer key (remote alarm is cancelled)
OFF: Remote alarm setting connected with
Buzzer key (remote alarm is not cancelled)
Note: Resolve the cause of the alarm, as the alarm itself is not deactivated by pressing the Buzzer key.
Table 6. In cases of high-limit temperature alarm or Err38
Remote Alarm setting
ON: Remote alarm setting not connected with
Buzzer key (remote alarm is cancelled)
OFF: Remote alarm setting connected with
Buzzer key (remote alarm is not cancelled)
Note: Close the outer door when Err38 is activated.
Table 7. In cases of Err01, Err11, Err12, Err18, or door alarm*
Remote Alarm setting
ON: Remote alarm setting not connected with
Buzzer key (remote alarm is cancelled)
OFF: Remote alarm setting connected with
Buzzer key (remote alarm is not cancelled)
*2: The remote alarm does not work for the door alarm.
Note: When Err01 is activated, connect the new CO
buzzer. In addition, when the optional MCO-21GC is installed and the gas supply is switched to the reserve
gas cylinder, press the Buzzer key and replace the gas cylinder.
Conditions
"Err18: UV Lamp Abnormal" is displayed
in the message display field.
"Warning: UV Bulb Life" is displayed in
the message display field.
"Err56: Communication Failure" is
displayed in the message display field.
"Status: Gas sensor initializing" is
displayed in the message display field.
Ring Back
When pressing
setting
the Buzzer key
ON
OFF
ON
OFF
Ring Back
When pressing
setting
the Buzzer key
ON
OFF
ON
OFF
Ring Back
When pressing
setting
the Buzzer key
ON
OFF
ON
OFF
Display
Buzzer from CO
incubator
2
When the Ring Back
set time passes
ON
OFF
OFF
(Alarm is not
ON
cancelled)
OFF
Buzzer from CO
incubator
2
When the Ring Back
set time passes
ON
ON
(Under continuation)
2
Buzzer from CO
incubator
2
When the Ring Back
set time passes
OFF
OFF
(Alarm is
(Alarm is already
cancelled)
cancelled)
gas cylinder and press the Buzzer key to stop the
2
Remote
Safety
Buzzer
alarm
operation
Intermittent
ON
tone
-----
-----
-----
-----
-----
-----
Remote Alarm
When pressing
When the Ring Back
the Buzzer key
set time passes
ON
ON
(Under continuation)
ON
OFF (Alarm is
not cancelled)
OFF
Remote Alarm
When pressing
When the Ring Back
the Buzzer key
set time passes
ON
ON
(Continue)
(Under continuation)
Remote Alarm
When pressing
When the Ring Back
the Buzzer key
set time passes
OFF
OFF
(Alarm is
(Alarm is already
2
cancelled*
)
cancelled*
-----
-----
-----
-----
2
)
65

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Mco-230aicuv

Tabla de contenido