Inserting The Batteries; Delivery Condition - Chauvin Arnoux C.A 10101 Guia De Inicio Rapido

Ocultar thumbs Ver también para C.A 10101:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

MEASUREMENT
The
symbol indicates the quality of the last electrode calibrated.
Slope value
Offset value
95% ≤ slope ≤ 105%
offset ≤ 19 mV
90% ≤ slope < 95%
19 mV < offset ≤ 38 mV
85% ≤ slope < 90%
38 mV < offset ≤ 58 mV
105% < slope < 85%
58 mV < offset
FRANCE
INTERNATIONAL
Chauvin Arnoux Group
Chauvin Arnoux Group
190, rue Championnet
Tél : +33 1 44 85 44 38
75876 PARIS Cedex 18
Fax : +33 1 46 27 95 69
Tél : +33 1 44 85 44 85
Fax : +33 1 46 27 73 89
Our international contacts
info@chauvin-arnoux.com
www.chauvin-arnoux.com/contacts
www.chauvin-arnoux.com
695716A00 - Ed. 1 - 11/2018
© Chauvin Arnoux - All rights reserved and reproduction prohibited
H O
2
State of the electrode
GETTING STARTED GUIDE TO THE C.A 10101 (GB)
USER'S MANUAL
Visit our web site to download the operating instructions for
your instrument and your electrodes:
www.chauvin-arnoux.com
Go to the Products tab and search on the name of your
instrument. When you have found the instrument, go to its
page. The operating instructions are on the right.
Download them.
Proceed in the same way for the pH and ORP electrodes.

INSERTING THE BATTERIES

DATA LOGGER TRANSFER SOFTWARE
Visit our web site to download the latest version of the ap-
plication software:
www.chauvin-arnoux.com
Go to the Support tab, then Download our software. Then
search on the name of your instrument.
Download the software, then install it on your PC.
KEYS
Key
Function
On/Off key (long press).
„ MEM: a short press to record the measure-
MEM
ment and the temperature.
REC
„ REC: a long press to start or stop recording.
„ pH/mV : a short press to select the type of
measurement (pH or ORP).
pH/mV
„ °C/°F: a long press to select the unit of
°C/°F
temperature.
Backlight.
„ CAL : a short press to look up the calibra-
CAL
tion parameters and start a calibration.
END
„ END: a long press to terminate the cali-
bration.
„ Un appui court pour régler la température
en MTC.
„ Un appui long pour choisir le set de solu-
tions tampons pendant l'étalonnage.

DELIVERY CONDITION

The C.A 10101 pH-meter is delivered in a carrying case with:
„ One XRGST1 combined pH electrode with integrated
temperature probe and storage reservoir.
„ Four AA or LR6 alkaline batteries.
„ One protective sheath fitted on the instrument.
„ Two ready-to-use NIST buffer solutions, pH 4.01 and 7.00.
„ Two beakers.
„ One USB to micro USB cable.
„ One wrist strap.
„ Getting started guides (one per language).
„ One certificate of verification.

Publicidad

Tabla de contenido
loading

Tabla de contenido