"A" AND "B" SUPPLIES
Output Voltage Range:
0V to 24VDC (0 ± 100mV)
Output Current Limit Range:
0 to 0.5A.
Constant Voltage Operation:
Voltage Regulation:
Load:
Line (108 – 132V):
Ripple Noise:
Recovery Time:
Temp. Coefficient
0°C to 40°C
Tracking Error:
No Load:
Full Load:
SPECIFICATIONS
≤0.01% + 3mV
≤0.01% + 3mV
≤2mVrms (5Hz to 1MHz)
≤100us typical.
≤300ppm/°C
≤0.2% + 20mV
≤0.2% + 100mV
Constant Current Operation:
Adjustable Current Limits:
Current Regulation:
Load:
Line (108 – 132V):
Ripple Current and Noise:
Metering ("A" & "B" only):
Voltmeter:
Range:
Accuracy:
Ampmeter:
Range:
Accuracy:
5
≤30mA to ≥500mA
≤0.2% + 6mA.
≤0.2% + 3mA.
≤3mA rms.
0 to 25V
≤2.5% of Full Scale
±2% + 2 digits
(1652)
0 to 600mA
≤2.5% of Full Scale
±2% + 2 digits
(1652)