Display
Code
EC,E21
Blinking (alternately)
Blinking (alternately)
Sensitivity adjustment weight error (Negative value)
The sensitivity adjustment weight is too light. Check around the pan. Check
the sensitivity adjustment mass value.
To return to weighing mode, press the
Overload error
A sample beyond the balance weighing capacity has been placed on the
pan. Remove the object from the pan.
Weighing pan error
The mass value is too light. Check that the weighing pan is installed
correctly. Set the weighing pan correctly. Calibrate the balance.
Sample mass error
The sample is too light to be stored as a sample mass for the counting mode
or percent mode. The sample cannot be used.
Unit mass error
The sample unit mass for the counting mode is too light. Storing and using
it for counting may cause a counting error. Add samples to reach the
specified number and press the PRINT key. Pressing the PRINT key
without adding samples will put the balance in counting mode. But, for
accurate counting, be sure to add samples.
Clock battery error
The clock backup battery has been depleted. Press any key and set the
time and date. Even if the clock backup battery is depleted, the clock and
calendar function works normally as long as the balance is powered with
the AC adapter. If this error appears frequently, contact your local dealer.
Power supply voltage fault
The voltage supplied from the AC adapter is abnormal.
Please check if the problem is the AC adapter supplied with the balance.
Repeatability error
The standard deviation (SD) of repeatability exceeded 50 digits. Review the
installation environment of the balance.
in repeatability display.
in minimum weighing value (reference value) display.
For details, refer to the separate instruction manual listed in"1.1. Detailed
Manuals".
Full memory
The number of stored weighing values has reached the upper limit. In order
to store new weighing values, it is necessary to delete data. For details, refer
to the separate instruction manual "1.1. Detailed Manuals".
Full memory
The stored sensitivity adjustment / calibration test history has reached 50
results. In order to store a new result, the oldest history will be deleted. For
details, refer to the separate instruction manual "1.1. Detailed Manuals".