9. Open the lamp housing using the door lock screw ①
and remove the lamp holder. (Fig. 26)
10. Remove the plastic block ② from the lamp holder (or the
exhausted lamp in case of replacement) by loosening
the two locking screws ③. (Fig. 27)
11. Insert the mercury bulb ④ (respect the polarity of the
bulb), tighten the locking screws and refit the lamp
holder inside the lamp housing. (Fig. 28)
•
Do not touch the bulb with bare hands.
12. Connect the cable from the external power supply to
the lamp housing. (Fig. 29)
Page 24