Troubleshooting
Byte
Bit
Description
21
1
SIMULATION_ACTIVE
2
SOFTWARE_UPDATE_INCOMPLETE
3
POWER_LOW
4
POWER_CRITICAL
5
FAULT_PREDICTION
6
ENVIRONMENTAL
7
OUTSIDE_LIMITS
22
0
REAL_TIME_CLOCK_NOT_VALID
1
WARNING_UCR_VOLTAGE
23
6
OTHER_FAULTS
24
3
NO_NETWORK_CONNECTION
4
SENSOR_VITALITY_POOR
5
SENSOR_CAL_INTERVAL_EXPIRED
6
SENSOR_WARMUP
58
Instructions for use
|
Dräger Polytron
Cause
Device is in simulation mode.
Firmware update active.
Firmware update failed.
Battery charge low.
Battery charge critical.
Device warning.
Device is operated outside the specified
temperature range.
Measured value too low.
No connection to the WirelessHART
work.
Error in sensor.
Serious device fault, various causes.
Incorrectly configured WirelessHART
work.
Vitality of the sensor is low.
Calibration interval expired.
EC sensor in warm-up phase 1.
®
6100 EC WL, Dräger Polytron
Remedy
Exit simulation mode via PolySoft.
Wait for firmware update (maximum
15 minutes).
A Restart the device.
B Have the gas detector checked by Dräger-
Service.
Change battery pack soon.
Change battery pack immediately.
®
A Check WirelessHART
B Check detail status with PolySoft.
In case of high temperatures due to solar radi-
ation, use weather protection set (accessory).
A Calibrate the sensor.
B Change the sensor.
®
net-
Connect to WirelessHART
Change the sensor.
Have the gas detector checked by DrägerSer-
vice.
®
®
net-
Check WirelessHART
network settings.
Change sensor soon.
Calibrate the sensor.
Wait for warm-up phase 1 to finish.
®
Repeater WirelessHART
connection.
®
network.
®