10. Figure 2-55 shows the completed cabling procedure inside the IND246 POWERCELL
2.4.4.5.
Cable Lengths
The terminal cannot be used with cable lengths exceeding those given in Table 2-5, or with more
than 12 PDX cells.
64084463
|
03
|
12/2017
enclosure.
Figure 2-55: Internal POWERCELL Cable Connections Completed
Table 2-5: Maximum PDX Cable Lengths
Total Cell-Cell Cable
(meters/feet)
130/426
METTLER TOLEDO IND246 and IND246 POWERCELL Installation Manual
Ground wires
connected at
grounding stud
Black ground
wire created in
step 4
Signal wires
twisted together
Home Run Cable
Number of
(meters/feet)
PDX Cells
300/984
< 12
2-29