Carl Zeiss
Fig. 7
Mounting the magnetic field
transmitter and the carrying handle
Fig. 8
Placing the reflected-light
illuminator onto the stand
16
Start-Up and Operation
3.1.2
Mounting the reflected-light illuminator onto
the stand
• Remove the tube, see section 3.5.1. If the
clamping screw is used to fix the tube to the
stand, replace it by the Allen set screw.
• Insert the magnetic field transmitter (Fig. 7/1)
for interlock as far as it will go into the stand at
the rear strut and screw it down.
• Push the carrying handle (Fig. 7/3) from the
back into the stand and fix it with two screws
(Fig. 7/2).
• Insert the reflected-light illuminator (Fig. 8/1), in
right angle and slightly inclined with its dovetail
mount into the stand (Fig. 8/4).
• Position the reflected-light illuminator (Fig. 8/1)
horizontally and turn it backwards with the
dovetail mount in the stand, align it to the
outer edges of the stand and tighten the set
screw (Fig. 8/3).
If the outer edges of the reflected-light
illuminator are not exactly aligned to
the stand, it may be that the reflected-
light illuminator cannot be switched on
due to the interlock feature causing
the interruption of the circuit to
protect against emergent LED light.
• Place the tube onto the stand and tighten the
clamping screw (Fig. 8/2), see section 3.5.1.
Primo Star / Primo Star iLED
Mounting the reflected-light
fluorescence illuminator
M60-2-0011 v 08/2008