Allow exteriors to remain wet for 1 minute.
⑥
Repeat the same procedure for the
⑦
lancing device (step
* After the pre-cleaning and disinfection procedure, the
control solution should be tested to confirm that the meter
works properly before using the meter. Control solution
tests should be performed with two different levels of
control solutions (Level A and B). Verify that the test results
are within the range printed on the test strip box. See
pages 24–26 for how to do a control solution test.
Note
If any of the following deterioration signs appear after pre-
cleaning or disinfecting, please stop using the system and
contact Customer Service at 1-800-429-5001 Mon–Sat, 9 am–
9 pm EST. At all other times, contact your healthcare
professional.
When the inscriptions on the exterior of the meter (or lancing
•
device) have been removed
When the color of the meter (or lancing device) has changed or
•
faded
When cracks or roughness develop on the meter (or lancing
•
device)
When a part of the segment on the meter display does not
•
flash
When control solution test results do not fall within the stated
•
range on the test strip box.
to step ⑥).
①
www.i-sens.com 43