Masimo® SET® SL
Patient Cables
Reusable
INDICATIONS
Masimo® SET® LNC SL-10 Patient Cables and Masimo SET LNOP® PC SL-12 Patient Cables have the same indications for use as
the applicable sensors. Please see the compatible sensor's directions for use for indications for use and prescribed information.
DESCRIPTION
Masimo SET SL Patient Cables are used with Masimo LNCS® and LNOP sensors and with Spacelabs Healthcare™ devices
containing Masimo SET Technology.
The SL cable was validated on the Ultraview SL Command Module with Masimo SET.
WARNINGS, CAUTIONS, AND NOTES
• Ensure the cable is physically intact, with no broken or frayed wires or damaged parts. Visually inspect the cable and
discard if cracks or discoloration are found.
• Carefully route patient cable to reduce the possibility of patient entanglement or strangulation.
• Do not attempt to reprocess, recondition or recycle Masimo sensors or patient cables as these processes may damage
the electrical components, potentially leading to patient harm.
• Masimo sensors and cables are designed for use with devices containing Masimo SET® oximetry or licensed to use
Masimo sensors.
• All sensors and cables are designed for use with specific monitors. Verify the compatibility of the monitor, cable and
sensor before use, otherwise degraded performance and/or patient injury can result.
• Failure to properly connect the sensor or the oximeter module to the cable will result in intermittent readings, inaccurate
results, or no reading.
• To avoid damage to the cable, always hold it by the connector rather than the cable when connecting or disconnecting
either end.
• Always refer to the oximeter module operator's manual for complete instructions or additional instructions.
• To prevent damage, do not soak or immerse the sensor or cable in any liquid solution. Do not attempt to sterilize the
cable or sensor.
• Caution: Replace the cable when a replace cable message is displayed, or when a low SIQ message is consistently
displayed while monitoring consecutive patients after completing the low SIQ troubleshooting steps identified in the
monitoring device operator's manual.
• Note: The cable is provided with X-Cal™ technology to minimize the risk of inaccurate readings and unanticipated loss
of patient monitoring. The cable will provide up to 17,520 hours of patient monitoring time. Replace the cable when the
patient monitoring time is exhausted. Some legacy monitoring devices do not distinguish between the cable and the
sensor. In these devices, a replace sensor or similar message could pertain either to the sensor or the cable.
INSTRUCTIONS
Attaching the SL Patient Cables
Attaching the SL Patient Cables to the Device
1. Orient the connector of the cable to mate with the patient cable connector on the Spacelabs Masimo SET device and
insert.
Attaching the LNC SL-10 Patient Cable to a LNCS Sensor
1. Insert the sensor connector completely into the SL patient cable connector (Fig. 1).
2. Completely close the protective cover (Fig. 2).
Attaching the PC SL-12 Patient Cable to a LNOP Sensor
1. Orient the connecting tab of the sensor so that the "shiny" contacts are pointed up. Mate the logo on the sensor to the
logo on the patient cable (Fig. 5).
2. Insert the tab of the sensor into the patient cable connector until there is a tactile or audible click of connection (Fig.
6). Gently tug on the patient cable connector to ensure a positive contact. Tape may be used to secure the cable to the
patient for ease of movement.
Disconnecting the SL Patient Cables
Disconnecting the LNC SL-10 Patient Cable from a LNCS Sensor
1. Lift the protective cover to gain access to the sensor connector (Fig. 3).
2. Pull firmly on the sensor connector to remove from the patient cable (Fig. 4).
Disconnecting the PC SL-12 Patient Cable from the LNOP Sensor
1. Place thumb and index finger on the buttons on either side of the patient cable connector (Fig. 7).
2. Press firmly on the buttons and pull to remove the sensor.
D IREC TIO N S FO R US E
Not made with natural rubber latex
LATEX
PCX-2108A
02/13
3
en
Non-Sterile
4886C-eIFU-0816