GB
3. Removing the syringe
• Empty the syringe completely.
• Push the stopping lever (4) upwards.
• Remove the syringe from the Minilab 201.
Please observe:
When the 25 ml and 50 ml syringe are used, the reusable adapter can be removed from the
syringe by turning counter clockwise, after removal from the Minilab 201.
• Avoid touching the outlet opening when removing the syringe (beware of potentially aggressive
solutions).
• When using viscous solutions, the tightness of the syringe is not guaranteed after several
pipettings.
5. MAINTENANCE
The Minilab 201 requires no maintenance. Potential clogging is removed best of all with warm
water or iso-propanol.
Please observe:
The Minilab 201 should not by dissembled by the user.
6. REMEDYING FAILURE
Possible failure
Syringe cannot be inserted.
Stopping lever cannot be
pushed down.
Filling slide cannot be moved
upwards.
Dosing volume is uncertain
7. TECHNICAL DATA
Weight: approx. 100 g
Accuracy and precision: cf. Table 2
Liquid: bi-distilled water
Reference temperature: 20° to 25° Celsius, ± 0.5° constant
Measurement: in accordance with German standard DIN 12650.
Table 2
Syringe volume
The data for syringe 0.5, 1.25 and 2.5 ml are based on a measurement carried out with a 200 µl
pipetting tip. The measurement for syringe 0.05 ml was performed with a 10 µl pipetting tip.
8. MINILAB 201 KIT AND ACCESSORIES
Kit:
• Minilab 201
• Instruction Manual
• 1 adapter for syringe 25 and 50 ml
Accesories
Table 3
Description
Syringe 0.05 ml
Syringe 0.50 ml
Syringe 1.25 ml
Syringe 2.50 ml
Syringe 5.00 ml
Syringe 12.50 ml
Syringe 25.00 ml
Syringe 50.00 ml
Syringe adapter
Reason
Filling slide has not been
pushed upwards up to the stop.
Syringe has not been placed
properly.
Stopping lever has not been
pushed down completely.
Adjusting wheel has not locked
into place.
(ml)
0.05
0.50
1.25
2.50
5.00
12.50
25.00
50.00
Inaccuracy
(%)
± 2.5
± 0.8
± 0.8
± 0.8
± 0.4
± 0.3
± 0.2
± 0.2
Cat.no.
5921
5922
5923
5924
5925
5926
5927
5928
5929
GB
Remedy
Push filling slide right up to the
stop
Remove syringe again, push
piston of the syringe right in,
push the filling slide right down
to the stop.
Press the stopping lever right
down to the stop
Lock adjusting wheel into place.
Imprecision
(%)
≤ 5.0
≤ 0.7
≤ 0.5
≤ 0.4
≤ 0.3
≤ 0.3
≤ 0.3
≤ 0.2
Number of pcs
per pack
100
100
100
100
100
100
25
25
1