Caution:
Only use the tester (MMT-7736) with the transmitter. Do not use any
other test plug. Other test plugs are not intended for use with the
transmitter, and will damage the transmitter and the tester.
Inspecting the transmitter connector pins
This image is an example of how the connector pins should look.
Look inside the connector opening of the transmitter to make sure that the connector pins
are not damaged or corroded. If the connector pins are damaged or corroded, the
transmitter cannot communicate with the charger, monitor, or pump. Contact the 24 Hour
HelpLine or your local representative. It may be time to replace your transmitter.
Also look for moisture inside the connector opening. If you see any moisture, allow the
transmitter to dry for at least one hour. Moisture inside the connector opening could
cause the transmitter to not work properly, and could cause corrosion and damage over
time.
i
connector opening
connector pins
housing
-8-
Correct tester has
gold on both sides
of connector.