Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment specified be-
low. The customer or the user of the transmitter should assure that it is used in such
an environment.
Immunity Test
Voltage dips, short in-
terruptions and voltage
variations on power
supply lines
IEC 61000-4-11
Power frequency
 (50/60 Hz) magnetic
field
IEC 61000-4-8
Note:
U
is the a.c. mains voltage prior to application of the test level.
T
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment specified be-
low. The customer or user of the transmitter should assure that it is used in such an
electromagnetic environment.
Immunity Test
Conducted RF
 IEC 61000-4-6
IEC 60601 Test Level
<5% U
(>95% dip in U
T
for 0.5 cycle
3 A/m
IEC 60601 Level
Compliance Lev-
el
3 V/m
 150 kHz to
80 MHz
Compliance
Level
)
Not applicable
T
400 A/m
Electromagnetic Environ-
ment Guidance
Not applicable
Not applicable
-24-
Electromagnetic Envi-
ronment - Guidance
Requirement does not ap-
ply to this battery pow-
ered device.
Power frequency magnet-
ic fields should be at lev-
els characteristic of a
typical location in a typical
domestic, commercial, or
hospital environment.