Advanced Bionics CI-5068 Manual Del Usuario página 23

Ocultar thumbs Ver también para CI-5068:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Cleaning and Maintenance
Remove the O-Ring prior to cleaning. Clean the
M Waterproof Battery with a damp cloth. Do not use
solvents. After exposure to moisture, wipe the battery
case with a soft, lint-free cloth and air dry overnight.
Inspect the O-Ring prior to each use, as it will need to be
changed periodically. If damage to the O-Ring is observed
or suspected, replace prior to resuming use.
If the M Waterproof Battery is dropped, inspect it for
evidence of damage or cracking. If any evidence of damage
is seen, the M Waterproof Battery should be replaced.
Ingress Protection (IP) Ratings
The following product configuration has an IP rating of 68
(complete protection against dust penetration; protection
against failure due to continuous immersion in water up
to 3 meters):
The Naída CI M90 sound processor inside the
M Waterproof Battery with a waterproof headpiece
and cable.
21

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ci-7323

Tabla de contenido