6.5 Conductivity measurement initial display

When switching on the CD-2004 («ON/OFF» key) the conductivity and temperature
measurement initial display appears.
Push «
/C/T/S» key to change measurements in conductivity, TDS and salinity.
Note: In order the measurements are precise, it's necessary to configure and
calibrate the conductivity meter.

6.5.1 Conductivity measurement: (CON)

In order this measurement is correct, you must use the configuration mode to set
up the following parameters:
Sensor constant.
Temperature compensation coefficient.
Conductivity sensor constant calibration.
6.5.2 Total Diluted Solids measurement: (TDS)
In order this measurement is correct, you must use the configuration mode to set
up the following parameters:
Sensor constant.
Temperature compensation coefficient.
«TDS transfering coefficient»
Conductivity sensor constant calibration
Coefficient calibration: «TDS transfering coefficient»
6.5.3 Salinity measurement: (SALT)
In order this measurement is correct, you must use the configuration mode to set
up the following parameters:
Sensor constant.
Conductivity sensor constant calibration
It's not necessary to insert a temperature compensation coefficient, as the equip-
ment performs this compensation automatically.

6.6 Suitable conductivity sensor selection

Sensor constant to measure in a conductivity range:
Conductivity
0.05 ... 2 µS/cm
2 ... 200 µS/cm
0.2 ... 2 mS/cm
2 ... 20 mS/cm
20 ... 200 mS/cm
Sensor constant to measure in a TDS range:
TDS
0 ... 1000 mg/L
1000 ... 10000 mg/L
1000 ... 19990 mg/L
Sensor constant to measure in a salinity range:
A 10 (cm
) constant sensor is normally used to measure salinity; if the salinity is
-1
lower than 1%, we can also use a 10 (cm
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso)
Resistivity (Ohm)
20M ... 500K
500K ... 5K
5K ... 500
500 ... 50
50 ... 5
Conductivity
0 ... 2 mS/cm
2 ... 20 mS/cm
20 ... 40 mS/cm
) constant sensor.
-1
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Pág..: 16
Constant (cm
)
-1
0.01 ; 0.1
0.1 ; 1
1
1 ; 5 ; 10
5; 10
Constant (cm
)
-1
1
1 ; 10
10