Limitations On Use - endolite Epirus Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

3 Function
The Epirus comprises an e-carbon toe and independent heel spring combined with resilient
multi-axial ankle movement, ankle function can be customised to the individuals requirements
via the use of interchangeable buffers.
Heel and toe springs are attached to the carrier using titanium and stainless Steel screws. The
foot is wrapped in a UHM PE sock which is in turn surrounded by a PU foot shell.
4 Maintenance
Maintenance must be carried out by competent personnel.
It is recommended that the following maintenance is carried out annually:
• Remove the foot shell and glide sock, check for damage or wear and replace if necessary.
• Check all screws for tightness, clean and reassemble as necessary.
• Visually check the heel and toe springs for signs of delamination or wear and replace if
necessary. Some surface damage may occur after a period of use, this does not affect the
function or strength of the foot.
• Remove the dome, check the condition of the plantar flexion buffer and replace if necessary
The wearer should be advised:
Any changes in performance of this device must be reported to the practitioner,
changes in performance may include:
• Increase in ankle stiffness
• Reduced ankle support (free movement)
• Any
unusual noises
The practitioner must also be informed of any changes in body weight and/or activity level.
Cleaning:
Use a damp cloth and mild soap to clean outside surfaces, do not use aggressive cleansers.

5 Limitations on use:

Intended life:
A local risk assessment should be carried out based upon activity and usage.
Lifting loads:
User weight and activity is governed by the stated limits.
Load carrying by the user should be based on a local risk assessment.
Environment:
Avoid exposing the Epirus to corrosive elements such as water, acids and other liquids. Also
avoid abrasive environments such as those containing sand for example as these may promote
premature wear.
Exclusively for use between -15˚C and 50˚C (5˚F to 122˚F).
It is recommended that only Endolite products be used in conjunction with the Epirus.
4
938276/5-0615

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ep22l1Ep30r8Ep22l1dEp30r8d

Tabla de contenido