Guidelines and manufacturer's declaration – electromagnetic immunity
The e-scope instrument is intended for use in the electromagnetic environment
specified below. The customer or user of the e-scope should ensure that it is used
in such an environment.
Immunity testing
Electrostatic
discharge (ESD)
IEC 61000-4-2
Fast transient
electrical distur-
bances / bursts
IEC 61000-4-4
Impulse voltage
IEC 61000-4-5
IEC 61000-4-11
Voltage dips, short
interruptions and
voltage variations
according to
IEC 61000-1-11
Magnetic field with
energy-efficient
rated frequencies
IEC 61000-4-8
Note: U
is the AC source. Mains voltage before the application of the test level.
T
30
IEC 60601 test
level
Con: ±8 kV
Air: ±2, 4, 8,
15 kV
5/50 ns, 100
kHz, ±2 kV
±0.5 kV voltage
Phase-to-phase
conductor
± 2 kV voltage
Line-to-earth
± 0.5 kV voltage
Phase-to-phase
conductor
± 2 kV voltage
Outer conductor
to earth
<0% UT 0.5
period at
0, 45, 90, 135,
180, 225,
270 and 315
degrees
0% UT 25/30
periods
Single phase: at
0 degrees
(50/60 Hz)
30A/m
50/60 Hz
Compliance
Con: ±8 kV
Air: ±2, 4, 8,
15 kV
Not applicable
Not applicable
Not applicabled
30A/m
50/60 Hz
Electromagnetic
environment -
guidance
Floors should be
wood, concrete or
ceramic tile.
If the floor is cover-
ed with synthetic
material, the relative
humidity must be at
least 30%.
The quality of the
supply voltage
should be that of a
typical commercial
or hospital environ-
ment.
The quality of the
supply voltage
should be that of a
typical commercial
or hospital environ-
ment.
The quality of the
supply voltage
should be that of a
typical commercial
or hospital environ-
ment.
Mains frequency
magnetic fields
should be at a level
characteristic of a
typical location in a
typical commercial
hospital environ-
ment.