1
1 Pt Adjust
Spray approximately 75 microns (3 mils) of powder onto a
2
flat, smooth metal sample.
3
Select 1 Pt Adjust then select Measure.
Take at least 3 uncured powder measurements with the
4
PosiTector PC. Results will be stored in the gage.
5
Select OK.
Cure the powder.
Gage may be powered off. It will retain the cal
NOTE:
6
measurements during a power cycle.
Measure cured thickness with a magnetic or eddy current
7
coating thickness gage. Record the results.
8
Select 1 Pt Adjust then select Adjust.
Adjust the displayed reading average lower (-) or higher (+)
to match the cured thickness value. Select OK.
1
2 Pt Adjust
Spray powder onto 2 flat, smooth metal samples. Place
approximately 40 microns (1.5 mils) on one and 75 microns
2
(3 mils) on the other.
3
Select 2 Pt Adjust then select Measure.
Take at least 3 uncured powder measurements on the thin
powder panel with the PosiTector PC. Results will be stored
4
in the gage. Select OK.
Take at least 3 uncured powder measurements on the thick
powder panel with the PosiTector PC. Results will be stored
5
in the gage. Select OK.
Cure the powder on both panels.
Gage may be powered off. It will retain the cal
NOTE:
6
measurements during a power cycle.
Measure cured thickness on both panels with a magnetic or
7
eddy current coating thickness gage. Record the results.
8
Select 2 Pt Adjust then select Adjust.
Adjust the displayed 1st point reading average lower (-) or
9
higher (+) to match the cured thickness values. Select OK.
Adjust the displayed 2nd point reading average lower (-) or
higher (+) to match the cured thickness values. Select OK.
Cal Settings Menu
Cal Settings Menu
8