TROUBLESHOOTING

Troubleshooting
Before contacting service personnel, refer to the following instructions to check and to correct
any failures for incorrect weight measurements.
The accuracy of the 450KL scale when calibrated at the factory is within +/- 0.1kg or ¼ pound.
For the most accurate weight measurement, use the scale on a hard level surface and stand in
the center of the platform. If the error seems excessive, check the following:
Symptom
Possible Cause
Bar does not "0"
Questionable
weight, or
balance
the scale does
not center
equally in the
trig loop when
the poise
weight is "0"
value
Bar does not
Bar sticking or
move freely
adjustment needed
Platform rocks
Base out of
excessively or
alignment
touches base
at any corner
Beam does not
Scale may have
move at all
improper steel rod
during
connection or poise
weighing
weight may have
process
been set at a higher
weight than patient's
actual weight.
Scale not
Scale out of
accurately
calibration.
measuring
weight
For further assistance please contact technical support toll free at 1 (800) 638-3722.
UM450KL REV20190116_Web
Corrective Action
The balance ball must be balanced such that the
pointer comes to rest in the center of the trig-loop
when both poise weights are set at "0" value.
Ensure that the upper and lower poise weights are
firmly against the shoulder of the beam.
Adjust the balance ball screw by turning the screw at
the left end of the bar until balanced.
Check if the pointer is touching the side of the trig
loop in its range of travel. Realignment of the beam
may be necessary if it is touching.
Ensure that the upper and lower poise weights are
firmly against the shoulder of the beam.
Adjust the balance ball screw by turning the screw at
the left end of the bar.
See instructions on next page.
Reset the weights to a lower weight.
Ensure the steel rod is connected properly, refer to
"Assembly Instructions" on page 5.
The scale can be recalibrated by loosening the lock
screw at the end of the long lever in the bottom of
the base with a ¾" hexagon socket wrench and
moving it outward (longer) in the elongated slot to
decrease the indicated weight or inward (shorter) to
increase the indicated weight. One sixteenth inch
movement of the screw will change the indicated
weight by approximately one-quarter pound per 100
pounds.