Safety and Quick Start Guide
7.3
DPI612 – HFP (1000 bar)
PRESSURIZED GASES AND FLUIDS ARE DANGEROUS. BEFORE CONNECTING
OR DISCONNECTING PRESSURE EQUIPMENT, SAFELY RELEASE ALL THE
PRESSURE.
ENSURE THAT THE SYSTEM IS BLED OF EXCESS AIR BEFORE OPERATION. (IF
THE ITEM UNDER TEST HAS A LARGE VOLUME, PRE-FILL WITH THE
PRESSURE FLUID TO ENSURE THAT TRAPPED AIR IS KEPT TO A MINIMUM).
To prevent damage to the pressure station, do not let dirt get into the
pressure mechanism. Before connecting equipment, make sure it is clean.
7.3.1
Release the pressure
1
Note: Fill the reservoir before attaching the device.
7.3.2
Filling, priming
Ice in the pressure mechanism can cause damage. If the temperature is less
than 4°C (39°F), drain all water from the DPI612 - HFP.
Page 16 of 20
2
Druck DPI612
[EN] English – 109M0089