FUNCTION KEY

"
"
Use " " key to turn backlight on and off.
"mils/µm"
Use "mils/µm" key to switch between mils and µm.
(1 mils = 25.4 µm)
"Zero"
1. Quickly push "Zero" key (no longer than 2 seconds) for substrate zeroing calibration.
2. Hold "Zero" key to calibrate frequent calibrating point.
3. Delete all calibration readings.
4. Delete MAX, MIN, and MAX-MIN readings.
"MAX/MIN"
1. Use "MAX/MIN" key to switch maximum, minimum, max-min, average, and total counts of data log
(MAX, MIN, MAX-MIN, AVG, and NO).
2. Capacity for data log is 255. Maximum, minimum, max-min, and average calculation will not be
refreshed after 255
3. When power is off, hold "MAX/MIN" key and pull the trigger to enter frequent calibrating point
setting.
"CAL"
1. When power is off, use "CAL" key and pull the trigger to enter Hi/Lo limit alarm setting.
2. When power is on, use "CAL" key for one point calibration.
3. In data logging mode and frequent calibrating point setting, use "CAL" to confirm and return to
operation