BGS technic 2197 Manual Del Usuario página 13

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

INSTRUCTION
Power on and off:
1. Keep the sensing tip of the meter away from any substrate or any magnetic field.
2. Pull the trigger to turn on power. When LCD shows "run" and
for use.
3. Auto Power Off (APO) function: Leave the gauge without operation for 30 seconds,
power turns off automatically.
Auto Mode and Fixed Mode:
1. The meter is set to auto mode, indicated as, which recognizes ferrous
and non-ferrous substrate automatically.
1. 2. If the substrate is fixed to ferrous or non-ferrous material, users
may use fixed mode. Hold button and pull the trigger to fix ferrous
mode. Hold "mils/µm" button and pull the trigger to fix non-ferrous mode. At the
mean time, will not be shown.
Hi/Lo Alarm Function
Hi/Lo alarm function is always on to alert users. When the measurement is over high limit, alarm
beeps 4 times; while the measurement is under low limit, alarm goes off continuously for 2.5
seconds. The Hi/Lo limit is defaulted 1200µm and 0µm.
User may set the limits for application. Hold "CAL" when power is off. Pull trigger to
power on and enter "SET Hi". Adjust Hi limit by using ▲ or ▼.
Confirm with "CAL" and enter "SET Lo". Adjust Lo limit by using ▲ or ▼. Confirm
again with "CAL" and the meter is ready for use.
Measuring:
1. Turn on the power.
2. Press the sensing tip to contact coated surface tightly. Pull the trigger and release immediately to
have single measurement.
the sensing tip from surface until
3. Substrate material will be indicated accordingly as "Ferr" or "Non-Ferr". If the substrate material
can not be recognized, "Ferr" and "Non-Ferr" are not shown.
4. Pull the trigger and hold it, continuous measuring will be performed. Reading is refreshed every
second. Release the trigger and wait for
the sensing tip from surface until
5. APO is inactivated during continuous measuring.
BGS technic KG
Bandwirkerstr. 3
D-42929 Wermelskirchen
sign appears when measurement is completed. DO NOT remove
sign is shown.
sign to complete the last measuring. DO NOT remove
sign is shown.
Tel.: 02196 720480
Fax.: 02196 7204820
mail@bgs-technic.de
, the meter is ready
A
N o n -
F e r r
www.bgstechnic.com
© BGS technic KG, Copying and further use not allowed
A L M
µm
H
M A X
F e r r
A L M H
i
µm
S E T
A L M
L o
µm
S E T

Publicidad

Tabla de contenido
loading

Tabla de contenido