Data Logging
The item automatic stored up to 255 readings
1.
Pull trigger
2.
Display shows "run"
Push "mils/um" button for 2 seconds
3.
(Display shows a Symbol)
4.
Main display shows the thickness
5.
Secondary display shows number of stored readings
Push „▲" or "▼" to change the stored readings
6.
7.
Push „CAL" button to left the data storage
If data storage is empty
8.
Display shows „no dAtA"
The unit exits the data storage
Delete data storage
Push „▲" or „▼" button until
9.
the display shows "CLr LoG".
Push „Cal" button to delete the data storage.
The unit exits the data storage.
CALIBRATION
Please have the standard coating plate and foil ready for one point calibration.
During the calibration the auto power off is extended to 2 minutes.
It is important that steps be observed during calibration.