*For part numbers of compatible oximeter accessories, refer to
www.resmed.com or contact your care provider. For information on how
to use these accessories, refer to the user guide that comes with these
accessories.
CAUTION
• In the US, Federal law restricts this device to sale by or on the order of a
physician.
• Only use the oximeter adapter as specified in this guide.
• S9 oximeter adapters are specifically designed for use with ResMed S9
devices.
• Air10 oximeter adapters are specifically designed for use with ResMed
AirSense 10 / AirCurve 10 / Lumis devices.
• Do not use liquids to clean any part of the oximeter adapter.
• Factors that may degrade pulse oximeter performance or affect the
accuracy of the measurement include the following: excessive ambient
light, excessive motion, electromagnetic interference, blood flow
restrictors (arterial catheters, blood pressure cuffs, infusing lines,
etc.), moisture in the sensor, improperly applied sensor, incorrect
sensor type, poor pulse quality, venous pulsations, anaemia or low
haemoglobin concentrations, cardiogreen or other intravascular dyes,
carboxyhaemoglobin, methaemoglobin, dysfunctional haemoglobin,
artificial nails or fingernail polish, or a sensor not at heart level.
Note: For any serious incidents that occur in relation to this device, these should
be reported to ResMed and the competent authority in your country.
Limited warranty
Refer to your device User Guide.
8