ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic emissions
The 04-610-001 is intended for use in the electromagnetic environment spec-
ified below. The customer or the user of the 04-610-001 should assure that it
is used in such an environment
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
IEC 61000-3-2
Not
Voltage fluctua-
Applicable
tions/flicker emis-
sions IEC 61000-3-3
20
Electromagnetic environment-
guidance
The 04-610-001uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment
The 04-610-001 is suitable for use in
all establishments other than domes-
tic and those directly connected to
the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
#91-032-610 04/15
©2010 Briggs Medical Service Company
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic immunity
The 04-610-001 is intended for use in the electromagnetic environment
specified below. The customer or the user of the 04-610-001 should assure
that it is used in such an environment.
IEC 60601test
IMMUNITY
Compliance
level
test
Electrostatic
discharge
± 6 kV contact
± 6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
Power fre-
quency
(50/60 Hz)
3 A/m
magnetic
field
IEC 61000-4-8
NOTE: U T is the a.c. mains voltage prior to application of the test level.
21
Electromagnetic
environment-
level
guidance
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
± 8 kV air
synthetic material,
the relative humid-
ity should be at
least 30%.
Power frequency
magnetic fields
should be at levels
characteristic of a
3 A/m
typical location in a
typical commercial
or hospital
environment.