Table 3
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture's declaration - electromagnetic immunity
The 04-610-001 is intended for use in the electromagnetic environment
specified below. The customer or the user of the 04-610-001 should assure
that it is used in such an environment.
IMMUNITY
IEC
Compliance
test
60601test
level
Radiated RF
3 V/m 80
IEC 61000-
MHz
3 V/m
4-3
to 2.5 GHz
22
Electromagnetic
environment-guidance
Portable and mobile RF com-
munications equipment should
be used no closer to any part of
the 04-610-001 , including cables,
than the recommended separa-
tion distance calculated from
the equation applicable to the
frequency of the transmitter.
Table 3 - Continued
For all ME EQUIPMENT and ME SYSTEMS
Radiated RF
3 V/m
IEC 61000-
80 MHz
3 V/m
4-3
to 2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
23
Recommended separation
distance:
d=1.2 P
80 MHz to 800 MHz
d=1.3 P
800 MHz to 2.5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m). Field
strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey a , should be less than
b
the compliance level
in each
frequency range. Interference
may occur in the vicinity of
equipment marked with the
following symbol: