en
|
Mechanical installation
®
If an individual Bluetooth
-pairing PIN has been assigned, a prompt
appears for entering the PIN.
✓ After successful connection, the blue LED lights up continuously.
®
Gas detector and Windows
end device with PolySoft are connected.
If required, use PolySoft to assign an individual Bluetooth
4.2.4
Establishing a Bluetooth
Prerequisite:
®
–
Bluetooth
interface on the gas detector and on the Windows
is activated.
–
A pairing was implemented between the gas detector and Windows
device.
● Connect via PolySoft.
Entering a PIN is not necessary.
✓ After successful connection, the blue LED lights up continuously.
®
Gas detector and Windows
end device with PolySoft are connected.
4.2.5
Connecting the ISA100 Wireless
®
● Connect Windows
end device to the respective management station of the
™
ISA100 Wireless
network.
The configuration is done with PolySoft.
5
Mechanical installation
WARNING
Risk of explosion!
In explosion-hazard areas, electrostatic discharge may cause a spark to ignite
on the antenna.
► Do not install the antenna in explosion-hazard areas subject to intense
electrostatic charging, for example, due to fast moving particles or
pneumatic transport of powders.
38
®
pairing PIN.
®
connection
®
end device
®
end
™
interface
Instructions for use
WARNING
Risk of explosion!
In explosion-hazard areas, electrostatic discharge may cause a spark to ignite
on the housing and sensing head.
► Do not install the gas detector in dust explosion-hazard areas subject to
intense electrostatic charging, for example, due to fast moving particles or
pneumatic transport of powders.
5.1
Installing the gas detector - installation
without mounting accessories
The gas meter is installed vertically on a flat surface with the antenna facing
upwards.
For installation with mounting accessories, see corresponding assembly
instructions.
Prerequisites:
–
A suitable installation location is selected (see "Safety information", section
"Mechanical installation", page 33).
CAUTION
Risk of explosion!
The gas detector is not tested in environments with increased oxygen content
(> 21 % O
).
2
► Do not use the gas detector in environments with increased oxygen content.
Install the gas detector:
● Install the gas detector using the drilling template (see appendix "Drilling
template") and M6 (1/4") hexagon socket head cap screws.
5.2
Installing the antenna
1. Remove the cover cap from the antenna connector (Fig. A 4).
2. Screw the antenna into the connector and tighten it by hand.
|
®
Dräger Polytron
6100 EC WL, Dräger Polytron
®
Repeater ISA