QUESTIONABLE RESULTS:
If the test result falls outside the range printed on the
test strip vial, check the "Troubleshooting Guide" locat-
ed in the "System Troubleshooting" section of the Own-
er's Manual and repeat the test.
Results that fall outside of the control range may indi-
cate:
• An error in how the test was performed
• Control solution bottle not shaken well
• Expired or contaminated control solution
• Control solution that is too warm or too cold
• Test strip deterioration
• Meter malfunction
• Test strips expired
Caution: Do not use the test strips or control
solution after the expiration date, as the results may
be inaccurate.
Contact Customer Care
If your test strips have been damaged or if you
continue to get out-of-range results, it means that the
system or control solution may not be working properly.
Do NOT use the system to test your blood glucose
level and contact Customer Care at 1.800.243.2636.
AutoCode
®
Owner's Manual
Blood Glucose Meter
34