v The control solution test is out of range.
Probable Cause
Error in performing
the test.
Control solution bottle Shake the control solution
not shaken well.
Expired or
contaminated
control solution.
Control solution that is Control solution should come
too warm or too cold.
Test strip
deterioration.
AutoCode
®
Actions
Read the instructions thoroughly
and repeat the test again.
bottle vigorously and repeat
the test again.
Check the expiration date and
the discard date of
control solution.
to room temperature before
testing.
Repeat the test with a new
test strip.