PROCESS DESCRIPTION
1.
General Pressure/Time graph for a process
NAME
P1 Instrument 134 °C
P2 Instrument 121 °C
NAME
P3 Liquid 121 °C
2.
3.
PARAMETER
Sterilisation temperature
Sterilisation time
Post treatment
Sterilisation temperature
Sterilisation time
Post treatment
PARAMETER
Sterilisation temperature
Sterilisation time
Post treatment
Steam is generated in the sterilizer through an external
water feed that may be preceded by a softening filter.
Condensation is led off to the drain.
PRE-TREATMENT (1)
The purpose of pre-treatment is to remove air from the
chamber and from the goods. Air prevents the requi-
site contact between the steam and micro-organisms
to be killed.
Pre-treatment consists of steam injection and evacua-
tion pulses in different processes depending on the
program selected. In addition, the humidification es -
sential to killing mainly takes place during the pre-
treatment stage.
STERILISATION (2)
Sterilisation takes place by maintaining the saturated
steam's relation (boiling point at a specific pressure).
The sterilisation phase is in progress during a pre-
programmed number of minutes at the preselected
temperature/pressure.
POST TREATMENT (3)
The purpose of post treatment is to reduce the mois-
ture content in the goods. During post treatment the
pressure drops to atmospheric pressure or lower –
depending on the selected program. Post treatment is
completed by air being drawn in until the chamber is
at atmospheric pressure.
STANDARD PROCESSES
AREA
5 - 180 min
5 - 180 min
5 - 180 min
5 - 180 min
LIQUID PROCESS (EXTRA)
AREA
5 - 180 min
Customised processes are possible.
STANDARD
134°C
5 min
5 min
121°C
20 min
5 min
STANDARD
121°C
20 min
Cooling to 95 °C
PROCESS DESCRIPTION