Operating Instructions Midwest E Pro / E Plus
66 04 206 D3686.201.01.02.02
7.3
Manual cleaning and disinfection
IMPORTANT: Manual conditioning is possible in exceptional cases if
Manual conditioning is possible in exceptional cases
the national/regional regulations are followed. The national/regional
regulations are to be checked before.
NOTICE! Condition immediately, or at the latest, one hour after
Immediately after treatment
treatment.
NOTICE! Never clean in an ultrasound bath!
NOTICE! Never immerse in disinfectant solution!
NOTICE! Only use Midwest® Plus Aerosol Spray.
Using the T1 spray
IMPORTANT: Use a soft, clean, and disinfected brush for cleaning.
Use soft brush
✔ Wear appropriate protective clothing.
Procedure _ Instrument without integrated spray _ without Alpro
✔ All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Only use
disinfectants with no protein-fixing properties.
7 Conditioning