2.3 Pipetting
Use the pipette only with tip attached!
Liquid aspiration
Set the required volume (see 2.2 Setting the volume) and attach the
appropriate size pipette tip(s) to the tip holder(s)
•
Press the volume control button
•
Hold the pipette upright and immerse approx. 3mm of the pipette
tip into the liquid.
•
Allow the volume control button
starting position.
•
Hold the pipette tip in the liquid for a further 2 seconds to avoid
sucking in air.
•
Remove the pipette tip from the liquid.
Dispensing fluids
•
Place the pipette tip against the inner wall of the tube or vessel, at
an angle of 30 – 45°.
•
Press the volume control button
reaches the first stop position.
•
Wait one second and press the volume control button
the second stop position to empty the pipette tip fully (blow out).
•
Keeping the volume control button
the pipette tip up the inner wall of the tube to remove any residual
liquid.
•
Allow the volume control button
starting position.
•
Press the tip ejector button
until the first stop position.
1
to return slowly back to the
1
slowly and evenly until it
1
pressed down, lightly drag
1
to return slowly back to the
1
to eject the pipette tip.
2
ErgoOne
ErgoOne
User Manual
User Manual
®
®
:
7
down to
1
7
7