INSTRUCTIONS FOR USE
Compatibility Specifications
The Allen 4-Post is compatible with:
Operating Room Table Compatibility
Note: Consult the corresponding IFU for the products mentioned in the above table.
Sterilization Instruction:
This device is not intended to be sterilized. Equipment damage may occur.
Cleaning and Disinfection Instruction:
WARNING:
Do not use bleach or products that contain bleach to clean the device. Injury or
equipment damage can occur.
After each use, clean the device with alcohol-based wipes.
Do not put the device into water. Equipment damage can occur.
Use a cloth and a quaternary ammonium disinfecting/cleaning solution to clean
and disinfect the device.
Read and follow the manufacturer's recommendation for low-level disinfection.
Read and follow the cleaning product's instructions. Use caution in areas where
liquid can get into the mechanism.
Wipe the device with a clean, dry cloth.
Make sure that the device is dry before you store it or use it again.
CAUTION: DO NOT IMMERSE PADS IN ANY LIQUID
CAUTION: DO NOT USE BLEACH OR PHENOLICS ON PADS
Document Number: D-720749
Version: C
Description
Allen Wing Set System – A-70200
Allen Foam Head Blocks – A-70862
Allen Foam Block Stand – A-70825
Allen Hip Cut-Out Wing set – A-70201
Allen Chest Wing set – A-70203
Allen Pediatric Wing set– A-70202
Allen Contour – A-70270
Replacement Long Straps – A-70208
The Allen 4-Post is compatible with the
following surgical table rail styles: US, EUR or
EU, UK, JP, DEN.