The products have been designed for a large number of sterilization
cycles.
The materials used in manufacture were selected accordingly. However
with every renewed preparation for use, thermal and chemical stresses
will result in ageing of the products. The maximum number of
sterilizations for optical fiber is 500 times.
6.1 Initial processing
6.1.1 Processing principles
It is only possible to carry out effective sterilization after the completion
of effective cleaning and disinfection. Please ensure that, as part of your
responsibility for the sterility of products during use, only sufficiently
validated equipment and product-specific procedures are used for
cleaning/disinfection and sterilization, and that the validated parameters
are adhered to during every cycle.
Please also observe the applicable legal requirements in your country as
well as the hygiene regulations of the hospital or clinic,especially with
regard to the additional requirements for the inactivation of prions.
6.1.2 Post-operative treatment
The post-operative treatment must be carried out immediately, no later
than 30 minutes after the completion of the operation. The steps are as
follows:
1. Remove the optical fiber from the Curing light Device,and rinse away
the dirt on the surface of product with pure water (or distilled water/
deionized water);
6