Preparation Before Cleaning; Automated Cleaning - B.A. International OPTIMA 10 BASES10 Modo De Empleo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 26
2. Dry the product with a clean, soft cloth and place it in a clean tray.
Notes
a) The water used here must be pure water, distilled water or deionized
water.

6.2 Preparation before cleaning

Steps
Tools: tray, soft brush, clean and dry soft cloth. Remove optical fiber
from main unit and put it into the clean tray.
Use a clean soft brush to carefully brush the optical fiber until the dirt
on surface is not visible. Then use soft cloth to dry the optical fiber and
put them into a clean tray. The cleaning agent can be pure water, distilled
water or deionized water.
6.3 Cleaning
The cleaning should be performed no later than 24 hours after the
operation.
The cleaning can be divided into automated cleaning and manual
cleaning. Automated cleaning is preferred if conditions permit.

6.3.1 Automated cleaning

•The cleaner is proved to be valid by CE certificationin accordance with
ENISO 15883.
▪There should be a flushing connector connected to the inner cavity of
the product.
▪The cleaning procedure is suitable for the product, and the irrigating
period is sufficient.
7

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido