GCE MediFlowTec Instrucciones De Uso página 7

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
Ensure the hose/humidifier is compatible with the flow outlet feature.
Push the hose onto the flow selector flow outlet/screw onto the humidi-
fier.
Ensure the hose/humidifier is well engages.
6.2.3. FLOW SETTING
Connect the flow selector to the source of medical gas.
Set requested medical gas flow by the handwheel.
Do not try to apply an excessive torque on the flow control knob when it
stops on the maximum flow position or in zero position.
Always ensure that the flow control knob has engaged and not placed
between two settings otherwise the flow selector will not deliver the cor-
rect flow of medical gas.
During use ensure that connected hose is not mechanically damaged,
throttled or broken, which would reduces or breaks supply of gas to the
patient.
The medical gas flow rate must be prescribed and administered by a clini-
cally trained user.
Fluctuating inlet pressure and ambient temperature can have influence
on accuracy of flow measurement.
6.3. AFTER USE
Before disconnecting flow selector from the medical pipe line system
terminal unit or medical regulator, set the handwheel on close position -
value "0".
Remove connected accessories from user outlet.
Disconnect flow selector from user outlet.
7. CLEANING
Remove general contamination with a soft cloth damped in oil free oxygen
compatible soap water & rinse with clean water.
Disinfections can be carried out with an alcohol-based solution (spray or
wipes).
If other cleaning solutions are used, check that they are not abrasive and
that they are compatible the product materials (including labels) and gas.
Do not use cleaning solutions containing ammonia!
Do not immerse in water or any liquid.
Do not expose to high temperature (such as autoclave).
To apply the cleaning solution, do not spray it, as the spray may enter the
inner parts of product and cause contamination or damage.
Do not use pressure water as it could damage or contaminate the
product.
7/84
EN

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mediflow iiMediflow +

Tabla de contenido