7. What do I do if...
... no weight is displayed with a load on the scale?
– Is the scale switched on?
– Check the batteries.
0. 0 00
...
does not appear before weighing?
– Press the Start key again - there must not be any load on the scale.
... one segment is illuminated permanently or not at all?
– That point has a fault. Inform your service agent.
... the display shows
– Battery voltage is dropping. You should replace the batteries in the next few days.
batt
...
appears in the display?
– The batteries are empty. Put in new batteries.
St0p
...
appears in the display?
– Maximum load has been exceeded.
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
Er:011:
...
appears in the display?
– The scale is overloaded or the load is unevenly distributed. Take the load off the scale
or distribute weight more evenly and restart scale.
Er:012:
...
appears in the display?
– The scale was switched on with too great a load on it. Remove load and restart.
8. Servicing/recalibration
Have the scales recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is required in all cases if one or more calibration seals have been dam-
aged or calibration counter contents does not match the number on the valid calibration
counter sticker (see "Checking that the contents of the calibration counter are correct" on
page 19). We recommend having your scales serviced by your local service agent before
having it recalibrated. Seca Customer Service will be pleased to give you further assist-
ance.