Q Q u u i i c c k k r r e e f f e e r r e e n n c c e e r r e e p p l l a a c c e e m m e e n n t t p p a a r r t t s s
These parts are available for purchase for the standard models, but subject to change. Call Stryker Customer Service: 1-
800-327-0770 for available parts and prices.
N N a a m m e e
Cover assembly, heel slope, END406 80 in.
Cover assembly, heel slope, END406, 84 in.
Cover assembly, heel slope, Nylon 80 in.
Cover assembly, heel slope, Nylon 84 in.
C C o o v v e e r r r r e e p p l l a a c c e e m m e e n n t t - - S S t t a a n n d d a a r r d d m m o o d d e e l l
C C A A U U T T I I O O N N - - Always take caution when you replace the support surface cover. The internal fire barrier contains fiberglass
fibers. Dust particles from the fibers may cause skin irritation.
T T o o o o l l s s r r e e q q u u i i r r e e d d : :
•
None
P P r r o o c c e e d d u u r r e e : :
1. Raise the bed height to the full up position.
2. Lower the fowler and gatch sections to the full down positions.
3. Unzip the cover. Start at the foot end patient right corner of the support surface and stop at the head end patient right
corner.
4. Fold the top of the cover to the patient's right side and then remove the foam crib assembly from the bed and set aside.
5. Remove and discard the cover.
6. Place the replacement cover, unzipped and open, on the bed with the black bottom cover on the litter and the top cover
folded over the patient's right side of the bed.
7. Place the foam crib assembly on top of the bottom part of the cover to make sure that the foam crib is aligned with the
cover.
8. Fold the top cover over the top of the foam crib assembly to make sure that the top cover is aligned with the foam crib
assembly.
9. Zip the cover to close. Start at the head end patient right corner and stop at the foot end patient right corner.
10. Verify proper operation before you return the product to service.
EN
12
N N u u m m b b e e r r
281505550012
281505550014
281505550018
281505550016
2815-009-001 Rev E.0