Draining

Drain the bath before moving or storing.
Ensure the fluid is at a safe handling temperature, ~40°C or lower. Wear protective
clothing and gloves.
Place a suitable receptacle underneath the drain.
Slowly turn the drain plug until flow is observed.
When the flow stops close the drain plug.
Draining
(GPD 10, GPD 20, GPD 28 and GPD 15D only)
Remove the drain hose from its
housing and insert the fitting into the
drain quick disconnect.
When the flow stops remove the
hose and reinsert it back into the
housing.

Accessories

Only use the accessories supplied by Fisher Scientific.

Drain Kit Accessory

Before filling the bath remove the
drain plug from the bath drain.
Wrap coupler with Teflon
®
equivalent, and install into bath drain.
Insert quick disconnect into hose.
To start draining snap quick
disconnect onto coupler.
To remove quick disconnect from
coupler and stop draining press
down on the disconnect's grey
release.
Figure 3-3 – GPD Bath Drain Hose
tape, or
Figure 3-4 – Drain Kit Accessory
Installation
Coupler
Quick Disconnect
Neoprene Hose (3/8ID x 5/8OD)
Isotemp
Chapter 3
|
3-7
|