Installation Overview - Hach TU5 Serie Instrucciones Para El Usuario

Módulo de limpieza automática
Ocultar thumbs Ver también para TU5 Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 70
Section 2 Installation
Pacemaker precautions. The instrument has an internal magnet. Keep the instrument a minimum of
5 cm (2 in.) from the user. A magnetic field can:
• Stop the stimulating pulses from the pacemaker that control the rhythm of the heart.
• Cause the pacemaker to supply the pulses irregularly.
• Cause the pacemaker to ignore the rhythm of the heart and supply pulses at a set interval.
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.

2.1 Installation overview

Figure 3
shows the installation overview with all of the clearances necessary.
Install the turbidimeter and do a leakage test on the system. Refer to the turbidimeter documentation.
Then, install the Automatic Cleaning Module.
6 English
W A R N I N G
C A U T I O N

Publicidad

Tabla de contenido
loading

Tabla de contenido