Temperature Measurement - Sonel CMP-3000 Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

5.10 Temperature measurement

To perform the measurement:
set the rotary switch at Temp ºC ºF ,
to change the unit, press MODE ,
place the adapter of the temperature probe in COM terminal
(black leg) and VΩ
place the temperature probe in the adapter, as shown in the
figure:
 thin pin of the probe (marked as +) fits to terminal +;
 thick pin of the probe (marked as K) fits to terminal –;
 reversed connection of the probe is mechanically
impossible ,
contact the head of the temperature probe to the device under
test. Maintain the contact of the probe head with the part of the
device under test, until the reading stabilizes.
read the measurement result on the display,
after completing the measurements, disconnect the probe from
the meter.
CAUTION!
Risk of burns. The temperature probe heats up, adapting to the
temperature of tested object.
58
Hz%CAPTemp (red leg):
CMP-3000 – USER MANUAL