Operating data
a)
measurement category according to IEC 61010-1 ...................... CAT IV 600 V (III 1000 V)
b)
type of insulation ........................................................................................... double, Class II
c)
housing type .............................................................................................. double-composite
d)
degree of housing protection acc. to EN 60529 ............................................................. IP40
e)
pollution degree .................................................................................................................... 2
f)
opening of measurement clamp ........................................................................ 48 mm (1.9")
g)
power supply of the meter .........................................................................6LR61 9 V battery
h)
diode test ..................................................................................... I = 0.3 mA, U
i)
continuity test ............................................................................ acoustic signal for R < 50 Ω
.................................................................................................... measuring current <0.5 mA
j)
indication for range exceeding .............................................................................OL symbol
k)
low battery indication ......................................................................................... symbol
l)
measurement rate ...................................................................2 measurements per second
m)
INRUSH function
▪ sampling time ............................................................. 37.5 Hz (RMS), 2.4576 MHz (clock)
▪ integration time......................................................................................................... 106 ms
n)
crest factor for 50/60 Hz and indications in the range of 5...100%
▪ range of 40 A and 400 A ................................................................................................. 3.0
▪ range of 1000 A............................................................................................................... 1.4
o)
range of the non-contact voltage detector ................................ 10...1000 V AC (50/60 Hz)
p)
response time for PEAK function .................................................................................. 1 ms
q)
temperature sensor .............................................................................. type K thermocouple
r)
input impedance ........................................................................................ 10 MΩ (V AC/DC)
s)
AC read-out ...............................................................................True RMS (A AC and V AC)
t)
AC bandwidth .................................................................................................... 50...1000 Hz
u)
display ............................................................................................... 5 digit LCD with backlit
................................................................................... 50.000 counts with function indicators
v)
dimensions ................................................................................................. 230 x 76 x 40 mm
w)
flexible clamp dimensions ............................................................................... 140 x 180 mm
x)
meter weight .................................................................................................................. 501 g
y)
meter weight (without batteries) .................................................................................... 456 g
z)
operating temperature ........................................................................................... +5..+40C
aa) operating humidity ........................................................................................................ < 80%
bb) storage temperature ............................................................................................. -20..+60C
cc) storage humidity ........................................................................................................... < 80%
dd) drop test ............................................................................................................................ 2 m
ee) maximum operating altitude ....................................................................................... 2000 m
ff)
Auto-Off function ............................................................................................ approx. 30 min
gg) compliance with the requirements of the following standards ............................EN 61010-1
........................................................................................ EN 61010-2-032, EN 61010-2-033
................................................................................................................................ EN 61326
hh) quality standard ...................................................................................................... ISO 9001
CMP-3000 – USER MANUAL
< 2.8 V DC
0
69